The high impedance, 350 MHz passive oscilloscope probe compatible with all RIGOL oscilloscopes.
The high impedance, 350 MHz passive oscilloscope probe compatible with all RIGOL oscilloscopes.
Features
Frequency Compensation
Before taking any measurement using a probe, first check the frequency compensation of the probe and adjust it to match the oscilloscope used. Generally, the oscilloscope provides a calibration signal output terminal at the front panel. Connect the probe (set the probe attenuation to 10X) to the signal output terminal to display a square waveform on the oscilloscope. Insert the adjustment tool into the compensation adjustment hole at the probe head and rotate it clockwise/counterclockwise until the properly compensated square waveform is displayed (flat-top square waveform).
Bandwidth | 1X: DC to 35 MHz, 10X: DC to 350 MHz |
Input Resistance | 1X: 1 MΩ ± 1% @ DC, 10X: 10 MΩ ± 1% @ DC |
Rise Time | 1X: 10 ns, 10X: 2.3 ns |
Input Capacitance | 1X: 50 pF ± 20 pF, 10X: 10 pF ± 5 pF |
Attenuation Ratio | 10:1 or 1:1 Switchable |
Max. Rated Input Voltage | 1X: 150 V RMSCAT II, 10X: 300 V RMSCAT II |
Compensation Range | 10 pF to 25 pF |
Weight (probe only) | About 40 g |
Cable Length | 120 ± 2 cm |
Probe Tip Diameter | 4.6 mm |
Temperature | Operating: 0 to +50°C, Nonoperating: -20 to +70°C |
Altitude | Altitude Operating: 2,000 m, Nonoperating: 15,000 m |
Max. Relative Humidity | 80% relative humidity for temperature up to +31°C, decreasing linearly to 40% relative humidity at +50°C |
Pollution Degree | Pollution Degree 2 |
Bandwidth | 1X: DC to 35 MHz, 10X: DC to 350 MHz |
Input Resistance | 1X: 1 MΩ ± 1% @ DC, 10X: 10 MΩ ± 1% @ DC |
Rise Time | 1X: 10 ns, 10X: 2.3 ns |
Input Capacitance | 1X: 50 pF ± 20 pF, 10X: 10 pF ± 5 pF |
Attenuation Ratio | 10:1 or 1:1 Switchable |
Max. Rated Input Voltage | 1X: 150 V RMSCAT II, 10X: 300 V RMSCAT II |
Compensation Range | 10 pF to 25 pF |
Weight (probe only) | About 40 g |
Cable Length | 120 ± 2 cm |
Probe Tip Diameter | 4.6 mm |
Temperature | Operating: 0 to +50°C, Nonoperating: -20 to +70°C |
Altitude | Altitude Operating: 2,000 m, Nonoperating: 15,000 m |
Max. Relative Humidity | 80% relative humidity for temperature up to +31°C, decreasing linearly to 40% relative humidity at +50°C |
Pollution Degree | Pollution Degree 2 |