The calibrator is designed to quickly diagnose ground fault and current leakage that is often caused by water in conduits and junction boxes. It features a backlit display, making it easier to operate in dark areas, along with isothermal brass blocks with screw terminals, which allows for the connection of bare thermocouple wires.
The calibrator is designed to quickly diagnose ground fault and current leakage that is often caused by water in conduits and junction boxes. It features a backlit display, making it easier to operate in dark areas, along with isothermal brass blocks with screw terminals, which allows for the connection of bare thermocouple wires.
Two Calibrators in One
Calibrate all your thermocouple instruments AND your milliamp loops with this unit. It has all the functions of a 14 type thermocouple calibrator and can source 0 to 24 milliamps or simulate a two wire transmitter like a milliamp calibrator. You may also perform stand alone calibrations of two wire transmitters by simultaneously supplying the thermocouple input signal while supplying the 24 volts to power up the transmitter and monitoring the milliamp output signal. Powers HART transmitters with built-in 250 ohm resistor simplifying hookups with HART communicators
Troubleshoot loop problems
Quickly diagnose ground fault and current leakage with the patented loop diagnostic technology
Calibrate directly in temperature (°C and °F)
Stop carrying around a millivolt source and thermocouple tables. The unit works with the thermocouples you use including types J, K, T, E, R, S, B, N, G, C, D, L (J-DIN), U (T-DIN) and P (Platinel II). Easily set any value quickly to within 0.1° with the adjustable "EZ-DIAL" plus store any three temperatures for instant recall with the EZ-CHECK™ switch. You may also calibrate from -13.000 to +80.000 mV. Connect your thermocouple and the unit measures the probe in degrees C or F. Secondary display shows the millivolt value corresponding to the thermocouple temperature.
Calibration Lab Accurate and Stable
The internal cold junction thermistor is accurate to ±0.05°C and is traceable to NIST. The sensor is thermally bonded to an isothermal mass including brass blocks with screw terminals for connection of bare thermocouple wires along with a miniature thermocouple connector for fast connections. An extremely stable voltage reference and low drift components make it more accurate than most other handheld and benchtop thermocouple calibrators
The unit is capable of meeting the requirements of an AMS 2750 Field Test Instrument when certified by an accredited laboratory.
Checkout all your 4 to 20 mA loops and instruments
With the built-in milliamp calibrator you can check, calibrate and measure all your current signal instruments in a 4 to 20 milliamp loop. It can be used at any access point in your loop. Source and Read 0.000 to 24.000 mA, Simulate a 2 Wire Transmitter or use the unit to simultaneously power your 2 Wire Transmitter and measure its output while simulating the sensor input
Easy to use in the plant, field or shop
With this unit you can check and calibrate all your thermocouple instruments, measure thermocouple Sensors, and calibrate 4 to 20 milliamp instruments. Automatic indication of connections on the display for simple hookups. Carry it without worry – it comes protected with a rubber boot and rugged, low profile switches. Easy to operate even in dark areas with the backlit display
Fast calibration with automatic output stepping
Choose between 2, 3, 5, 11 and 21 steps to automatically increment the output in 100%, 50%, 25%, 10% or 5% of span. Select the step time to match your system from 5, 6, 7, 8, 9, 10, 15, 20, 25, 30 and 60 seconds
Temperature | Operating: -10 to 140°F (-25 to 60°C) Drift: ±0.01% of span outside of 73°C±18°F (23°C±10°C) |
Relative Humidity Range | 10% ≤RH ≤90% (32 to 95°F [0 to 35 °C]), non-condensing 10% ≤RH ≤70% (95 to 140°F [35 to 60 °C]), non-condensing |
Voltage | 60 V rms between all milliamp functions/Read V DC and Source V DC/Thermocouple |
Mode Rejection | Normal: 50/60 Hz, 50 dB Common: 50/60 Hz, 120 dB |
Display | High contrast graphic liquid crystal display. LED backlighting for use in low lit areas. |
Battery | 4 x AA Alkaline 1.5 V (LR6) Life: 25 hours RTD, 8 hours milliamp Low battery: Low battery indication with nominal 1 hour of operation left Optional NiMh rechargeable battery kit: Charger, four NiMh batteries |
Dimensions | 5.63 x 3 x 1.60" (143 x 76 x 41 mm) |
Weight | 12.1 oz (0.34 kg) |
Temperature | Operating: -10 to 140°F (-25 to 60°C) Drift: ±0.01% of span outside of 73°C±18°F (23°C±10°C) |
Relative Humidity Range | 10% ≤RH ≤90% (32 to 95°F [0 to 35 °C]), non-condensing 10% ≤RH ≤70% (95 to 140°F [35 to 60 °C]), non-condensing |
Voltage | 60 V rms between all milliamp functions/Read V DC and Source V DC/Thermocouple |
Mode Rejection | Normal: 50/60 Hz, 50 dB Common: 50/60 Hz, 120 dB |
Display | High contrast graphic liquid crystal display. LED backlighting for use in low lit areas. |
Battery | 4 x AA Alkaline 1.5 V (LR6) Life: 25 hours RTD, 8 hours milliamp Low battery: Low battery indication with nominal 1 hour of operation left Optional NiMh rechargeable battery kit: Charger, four NiMh batteries |
Dimensions | 5.63 x 3 x 1.60" (143 x 76 x 41 mm) |
Weight | 12.1 oz (0.34 kg) |