Ideal for emission monitoring of combustion and industrial processes, this combustion/emission analyzer measures oxygen, carbon monoxide, differential pressure, and temperature. In addition, this analyzer features a high-energy lithium-ion battery for up 20 hours of operation time and calculated carbon dioxide measurement.
Ideal for emission monitoring of combustion and industrial processes, this combustion/emission analyzer measures oxygen, carbon monoxide, differential pressure, and temperature. In addition, this analyzer features a high-energy lithium-ion battery for up 20 hours of operation time and calculated carbon dioxide measurement.
Cost-effective and user-friendly, this combustion/emission analyzer features a high-resolution display for providing more information with detailed graphics. Additionally, this analyzer features an illuminated condensate trap and water-stop and WLAN network connectivity.
Features
Applications
Oxygen (Long Life) | Measuring range: 0 to 21% Resolution: 0.1% Accuracy: ±0.2 volume % abs |
Carbon Monoxide (H2 Compensated) | Measuring range: 0 to 10,000/20,000 ppm Resolution: 1 ppm Accuracy: ±10 ppm or 5% reading <4000 ppm, 10% reading >4000 ppm |
Temperature | Stack: Measuring range: 32 to 1472°F (0 to 800°C) with stainless steel Resolution: 1°F (-17.22°C) Accuracy: ±4 to <392°F (-15.55 to <200°C), 1% reading >392°F (200°C) Flue gas: Measuring range: 32 to 2012°F (0 to 1100°C) with inconel Resolution: 1°F (-17.22°C) Accuracy: ±4 to <392°F (-15.55 to <200°C), 1% reading >392°F (200°C) Primary/ambient air: Measuring range: 32 to 212°F (0 to 100°C) Resolution: 1°F (-17.22°C) Accuracy: ± 2°F (-16.66°C) Differential: Measuring range: Up to 2012°F (0 to 1100°C) Resolution: 1°F (-17.22°C) Accuracy: ±4 to <392°F (-15.55 to <200°C), 1% reading >392°F (200°C) |
Stack Draft | Measuring range: ±40 in H2O (100 hPa) Resolution: 1 Pa Accuracy: ±0.02 in H2O or 1% reading |
Differential Pressure | Measuring range: ±80 in H2O (200 hPa) Resolution: 1 Pa Accuracy: ±0.02 in H2O or 1% reading |
Gas Flow Velocity Measurement | Measuring range: 3 to 100 m/s (using pitot tube) Resolution: 0.1 m/s |
Calculated Values | Carbon dioxide: 0 to CO2 maximum Efficiency: 0 to 120% Excess air: 0 to 99.9 Heat losses qA: 0 to 99.9% Air ratio (lambda): 1 to 9.99 CO/CO2 ratio: 0 to 10 |
Maximum Suction Range Gas Pump/Typical Gas Flow | 150 hPa/50 l/h |
Internal Memory | 32,000 data sets |
Data Transmission/Interfaces | USB, Bluetooth, WLAN/USB, Bluetooth, WLAN, IRDA, SD card |
Display | 4" (10.16 cm) color display |
Ambient Conditions | 95% relative humidity, non-condensing |
Protection Class | IP30 |
Power | Internal battery pack: Li-Ion, 20 hours operation time Grid power supply: 100 to 240 V, 5 V DC, 1200 mA |
Dimensions | 4.3 x 8.8 x 2.04" (244 x 113 x 54 mm) |
Weight | 1.65 lbs (750 g) |
Click on a category to view a selection of compatible accessories with the MRU Ampro Plus Combustion/Emission Analyzer, O2/CO.
Oxygen (Long Life) | Measuring range: 0 to 21% Resolution: 0.1% Accuracy: ±0.2 volume % abs |
Carbon Monoxide (H2 Compensated) | Measuring range: 0 to 10,000/20,000 ppm Resolution: 1 ppm Accuracy: ±10 ppm or 5% reading <4000 ppm, 10% reading >4000 ppm |
Temperature | Stack: Measuring range: 32 to 1472°F (0 to 800°C) with stainless steel Resolution: 1°F (-17.22°C) Accuracy: ±4 to <392°F (-15.55 to <200°C), 1% reading >392°F (200°C) Flue gas: Measuring range: 32 to 2012°F (0 to 1100°C) with inconel Resolution: 1°F (-17.22°C) Accuracy: ±4 to <392°F (-15.55 to <200°C), 1% reading >392°F (200°C) Primary/ambient air: Measuring range: 32 to 212°F (0 to 100°C) Resolution: 1°F (-17.22°C) Accuracy: ± 2°F (-16.66°C) Differential: Measuring range: Up to 2012°F (0 to 1100°C) Resolution: 1°F (-17.22°C) Accuracy: ±4 to <392°F (-15.55 to <200°C), 1% reading >392°F (200°C) |
Stack Draft | Measuring range: ±40 in H2O (100 hPa) Resolution: 1 Pa Accuracy: ±0.02 in H2O or 1% reading |
Differential Pressure | Measuring range: ±80 in H2O (200 hPa) Resolution: 1 Pa Accuracy: ±0.02 in H2O or 1% reading |
Gas Flow Velocity Measurement | Measuring range: 3 to 100 m/s (using pitot tube) Resolution: 0.1 m/s |
Calculated Values | Carbon dioxide: 0 to CO2 maximum Efficiency: 0 to 120% Excess air: 0 to 99.9 Heat losses qA: 0 to 99.9% Air ratio (lambda): 1 to 9.99 CO/CO2 ratio: 0 to 10 |
Maximum Suction Range Gas Pump/Typical Gas Flow | 150 hPa/50 l/h |
Internal Memory | 32,000 data sets |
Data Transmission/Interfaces | USB, Bluetooth, WLAN/USB, Bluetooth, WLAN, IRDA, SD card |
Display | 4" (10.16 cm) color display |
Ambient Conditions | 95% relative humidity, non-condensing |
Protection Class | IP30 |
Power | Internal battery pack: Li-Ion, 20 hours operation time Grid power supply: 100 to 240 V, 5 V DC, 1200 mA |
Dimensions | 4.3 x 8.8 x 2.04" (244 x 113 x 54 mm) |
Weight | 1.65 lbs (750 g) |
Click on a category to view a selection of compatible accessories with the MRU Ampro Plus Combustion/Emission Analyzer, O2/CO.