With up to 0.025% of reading accuracy of differential pressure, this reference recorder has a 300 psi range. Used to calibrate anywhere and accurately to replace a deadweight tester. Additionally, it is rugged enough to sit in 3' of water and has a BARO module option.
With up to 0.025% of reading accuracy of differential pressure, this reference recorder has a 300 psi range. Used to calibrate anywhere and accurately to replace a deadweight tester. Additionally, it is rugged enough to sit in 3' of water and has a BARO module option.
Perform complete tests and calibration anywhere.
Features
Applications
Upper Module | Pressure Range: 300 psi Accuracy: 0 to 30% of range: ±(0.0075% of full scale) 30 to 110% of range: ± (0.025% of reading) Vacuum: ± (0.06% of full scale) ±1 LSD |
Barometric Reference (BARO) | Accuracy: ± 0.00725 psi, ± 0.5 mbar Range: 10.153 to 15.954 psiA, 700.0 to 1100.0 mbarA Units and Resolution: psi. . . . . . . . . . . . . . . . . . 0.001 inHg. . . . . . . . . . . . . . . . 0.001 mmHg. . . . . . . . . . . . . 0.01 mbar. . . . . . . . . . . . . . . 0.1 Pressure Connection: Cylindrical sensor fitting of 0.23" (5.8 mm) OD. A flexible 3/16" (4.8 mm) ID tube is recommended to connect for for calibration Mounting: Secured using a 3/8" 4-40 plastic screw |
Sensor | Wetted materials: (wrench tight) 316 stainless steel, (finger tight) 316 stainless steel Viton (internal o-ring) Diaphragm seal fluid: silicone oil Connection: crystal CPF female |
Display | Screen: 255 x 160 pixel graphical display Display rate: 4 readings/second (standard) up to 10 readings/second (recording) |
Digital Interface | Mini-USB |
Data Logging | Capacity: approximately 1,000,000 data points Storage type: Non-volatile flash memory Fastest interval: 10 per second Slowest interval: 1 per hour |
Enclosure | Rating: IP66 and IP67 Housing: impact resistant injection molded Keypad and labels: UV resistant polyester Mounting: M4 x 0.7 [0.31" (8 mm)] deep threaded insert mounting locations |
Power | 4 x AA: 200 hours, typical Ultra low power: Up to 60 days, typical |
Weight | 1.5 lb (680 g) |
Upper Module | Pressure Range: 300 psi Accuracy: 0 to 30% of range: ±(0.0075% of full scale) 30 to 110% of range: ± (0.025% of reading) Vacuum: ± (0.06% of full scale) ±1 LSD |
Barometric Reference (BARO) | Accuracy: ± 0.00725 psi, ± 0.5 mbar Range: 10.153 to 15.954 psiA, 700.0 to 1100.0 mbarA Units and Resolution: psi. . . . . . . . . . . . . . . . . . 0.001 inHg. . . . . . . . . . . . . . . . 0.001 mmHg. . . . . . . . . . . . . 0.01 mbar. . . . . . . . . . . . . . . 0.1 Pressure Connection: Cylindrical sensor fitting of 0.23" (5.8 mm) OD. A flexible 3/16" (4.8 mm) ID tube is recommended to connect for for calibration Mounting: Secured using a 3/8" 4-40 plastic screw |
Sensor | Wetted materials: (wrench tight) 316 stainless steel, (finger tight) 316 stainless steel Viton (internal o-ring) Diaphragm seal fluid: silicone oil Connection: crystal CPF female |
Display | Screen: 255 x 160 pixel graphical display Display rate: 4 readings/second (standard) up to 10 readings/second (recording) |
Digital Interface | Mini-USB |
Data Logging | Capacity: approximately 1,000,000 data points Storage type: Non-volatile flash memory Fastest interval: 10 per second Slowest interval: 1 per hour |
Enclosure | Rating: IP66 and IP67 Housing: impact resistant injection molded Keypad and labels: UV resistant polyester Mounting: M4 x 0.7 [0.31" (8 mm)] deep threaded insert mounting locations |
Power | 4 x AA: 200 hours, typical Ultra low power: Up to 60 days, typical |
Weight | 1.5 lb (680 g) |