Combined pressure and mA loop measurements to be fully temperature compensated from -4 to 122°F (-20 to 50°C), this single pressure calibrator has the BARO option to toggle between the gauge and absolute pressure. Reliable accuracy in measuring pressure, current, voltage, or temperature.
Combined pressure and mA loop measurements to be fully temperature compensated from -4 to 122°F (-20 to 50°C), this single pressure calibrator has the BARO option to toggle between the gauge and absolute pressure. Reliable accuracy in measuring pressure, current, voltage, or temperature.
Used in a broad range of applications from simple tool type jobs to complex calibration jobs in custody transfer systems.
Features
Applications
Pressure Range | 300 psi |
Accuracy | 0 to 30% of range: ±(0.015% of full scale) 30 to 110% of range: ±(0.050% of reading) Vacuum : ±(0.05% of full scale) |
Pressure Sensor | Wetted materials: (wrench tight) 316 stainless steel, (finger tight) 316 stainless steel, and Viton (internal o-ring) Diaphragm seal fluid: silicone oil Connection: crystal CPF female |
Barometric Reference (BARO) | Accuracy: ±0.00725 psi, ±0.5 mbar Range: 10.153 to 15.954 psiA, 700.0 to 1100.0 mbarA |
Current and Voltage Measurement | Connection: 4 mm jacks Maximum voltage: 45 VDC Current (mA) input accuracy: ±(0.015% of rdg + 0.002 mA), mA range: 0 to 55 mA Current (mA) output accuracy: ±(0.015 of rdg + 0.002 mA), range: 0 to 25 mA Voltage (VDC) Input accuracy: ±(0.015 % of rdg + 2 mV), range: 0 to 30 VDC Loop power accuracy: ±(0.015% of rdg) + 0.02 Ω, Range: 0 to 400 Ω |
Enclosure | Rating: IP65 Housing: machined aluminum Keypad and labels: UV resistant silicone |
Display | 320 x 240 pixel graphical display |
Temperature Range | -4 to 122°F (-20 to 50°C) |
Power | Battery Life: >12 hours non-sourcing >8 hours when sourcing 12 mA |
Pressure Range | 300 psi |
Accuracy | 0 to 30% of range: ±(0.015% of full scale) 30 to 110% of range: ±(0.050% of reading) Vacuum : ±(0.05% of full scale) |
Pressure Sensor | Wetted materials: (wrench tight) 316 stainless steel, (finger tight) 316 stainless steel, and Viton (internal o-ring) Diaphragm seal fluid: silicone oil Connection: crystal CPF female |
Barometric Reference (BARO) | Accuracy: ±0.00725 psi, ±0.5 mbar Range: 10.153 to 15.954 psiA, 700.0 to 1100.0 mbarA |
Current and Voltage Measurement | Connection: 4 mm jacks Maximum voltage: 45 VDC Current (mA) input accuracy: ±(0.015% of rdg + 0.002 mA), mA range: 0 to 55 mA Current (mA) output accuracy: ±(0.015 of rdg + 0.002 mA), range: 0 to 25 mA Voltage (VDC) Input accuracy: ±(0.015 % of rdg + 2 mV), range: 0 to 30 VDC Loop power accuracy: ±(0.015% of rdg) + 0.02 Ω, Range: 0 to 400 Ω |
Enclosure | Rating: IP65 Housing: machined aluminum Keypad and labels: UV resistant silicone |
Display | 320 x 240 pixel graphical display |
Temperature Range | -4 to 122°F (-20 to 50°C) |
Power | Battery Life: >12 hours non-sourcing >8 hours when sourcing 12 mA |