Designed to meet the most stringent demands in industry and electrical contracting, they also meet the latest safety and performance standards. Features a measurement range of 0.1 to 150A and a large jaw opening accommodates conductors up to 250kcmil.
Designed to meet the most stringent demands in industry and electrical contracting, they also meet the latest safety and performance standards. Features a measurement range of 0.1 to 150A and a large jaw opening accommodates conductors up to 250kcmil.
The MN353 is a high accuracy and low phase shift AC current probe with voltage output. It is designed for use with DMMs or current measuring instruments with voltage inputs. The voltage output is proportional to the current measured (through the jaws). The MN353 has an integral 5 ft (1.5m) lead terminated by shrouded banana plugs.
Features
Applications
Nominal Range | 150A |
Measurement Range | 0.1 to 150A |
Transformation Ratio | Voltage output |
Output Signal | 10mV/A (1V at 100A AC) |
Accuracy (150A Range) | 0.1 to 1A: ≤3% of Reading ± 200µV 1 to 80A: 1% of Reading ± 0.2mV 80 to 150A: 1.5% of Reading ± 0.2mV |
Phase Shift (150A Range) | 1 to 20A: ≤3.0° 20 to 100A: ≤2.0° 100 to 150A: ≤2.5° |
Overload | 240A for 10 min ON, 30 min OFF |
Frequency Range | 40 to 10kHz |
Limit Operating | 200A permanently to 1kHz Derating above 3kHz: 200A x (1/0.333F), F in kHz to 1kHz |
Crest Factor | 3 at 150A Peak with an error (due to CF) of 3% |
Working/Common Mode Voltage | 600Vrms |
Output Termination | 5ft Lead |
Nominal Range | 150A |
Measurement Range | 0.1 to 150A |
Transformation Ratio | Voltage output |
Output Signal | 10mV/A (1V at 100A AC) |
Accuracy (150A Range) | 0.1 to 1A: ≤3% of Reading ± 200µV 1 to 80A: 1% of Reading ± 0.2mV 80 to 150A: 1.5% of Reading ± 0.2mV |
Phase Shift (150A Range) | 1 to 20A: ≤3.0° 20 to 100A: ≤2.0° 100 to 150A: ≤2.5° |
Overload | 240A for 10 min ON, 30 min OFF |
Frequency Range | 40 to 10kHz |
Limit Operating | 200A permanently to 1kHz Derating above 3kHz: 200A x (1/0.333F), F in kHz to 1kHz |
Crest Factor | 3 at 150A Peak with an error (due to CF) of 3% |
Working/Common Mode Voltage | 600Vrms |
Output Termination | 5ft Lead |